SONG Bo,ZHANG Shi-hong,YU De-zhao,et al.The Approximate True Turbidity Theory and Its Application in Low Turbidity Water Measurement[J].China Water & Wastewater,2020,36(14):146-152.
近真浊度理论及其在低浊度测量中的应用
- Title:
- The Approximate True Turbidity Theory and Its Application in Low Turbidity Water Measurement
- Keywords:
- turbidity; apparent turbidity; approximate true turbidity; low turbidity water sample; zero turbidity; particle
- 摘要:
- 对于低浊度水样(浊度<0.1 NTU),不同仪器间的检测结果差异明显,因此如何获得更客观的检测数据尤为重要。为此,首次提出了近真浊度理论,并给出了其测定方法。近真浊度也称为近似真实浊度,是指一种无限趋近于仅由悬浮颗粒物及胶体所引起的浊度,即无限趋近于浊度真实值,这些颗粒物和胶体包括泥砂、黏土、藻类、有机物质、微生物有机体等。通过在表观浊度值中扣除现有的技术手段制备出的无限接近零浊度的实验室用纯水浊度值即仪器的零示值误差,就可以得到近真浊度值。近真浊度和现行的浊度概念的最大的区别在于,近真浊度去除了由光源、仪器光路、检测器种类、系统杂散光、信号值计算方法、校准验证方法等方面产生的零示值误差,去除了零示值误差的结果会无限接近于浊度真实值。近真浊度这一概念的提出可以有效解决低浊度领域的精确测量问题。
- Abstract:
- For low turbidity water samples(turbidity<0.1 NTU), the test results of different instruments are significantly different, so how to obtain more objective test data is particularly important. Therefore, this paper first presented a new theory of the approximate true turbidity, and introduced its determination method. The approximate ture turbidity,also known as approaching to the true turbidity, refers to an infinite approach to turbidity caused only by suspended particles and colloids, that is, infinitely close to the true value of turbidity. These particles and colloids include: sand, clay, algae, organic matter, microbial organisms and so on. The approximate true turbidity can be obtained by subtracting the turbidity value of laboratory pure water (zero indication error) from the apparent turbidity. The biggest difference between approximate true turbidity and current turbidity concept is that the approximate true turbidity eliminates error of indication caused by light source, instrument optical path, detector type, system stray light, signal value calculation method, calibration verification method and other aspects. The result will be close to the true value of turbidity by eliminating the zero indication error. The concept of approximate true turbidity can effectively solve the problems of accurate measurement in the field of low turbidity.
相似文献/References:
[1]孙书博,谢益佳,黄廷林,等.循环造粒流化床处理水厂排泥水的试验研究[J].中国给水排水,2020,36(7):12.
[2]张安嫒.滤料粒径对地下水中铁、锰、氨氮、浊度去除效果的影响[J].中国给水排水,2020,36(7):47.
[3]李婷,董志峰,张松芳,等.东江水源和本地水库水源原水混用的给水厂运行实践[J].中国给水排水,2020,36(22):23.
LI Ting,DONG Zhi-feng,ZHANG Song-fang,et al.Operation Practice of Mixing Raw Water from Dongjiang River and Local Reservoir in a Waterworks[J].China Water & Wastewater,2020,36(14):23.
[4]彭进湖,何孙胃,陈丽珠,等.细砂滤料的过滤性能及运行效果[J].中国给水排水,2021,37(1):40.
PENG Jin-hu,HE Sun-wei,CHEN Li-zhu,et al.Filtration Performance and Operation Effect of Fine Sand Filter[J].China Water & Wastewater,2021,37(14):40.
[5]施凯,陈志伟,彭秀华,等.冰浆清洗技术对小口径球墨铸铁给水管道的清洗效果[J].中国给水排水,2021,37(3):69.
SHI Kai,CHEN Zhi-wei,PENG Xiu-hua,et al.Cleaning Effect of Ice Slurry Cleaning Technology on Small Diameter Ductile Iron Water Supply Pipeline[J].China Water & Wastewater,2021,37(14):69.
[6]袁文璟,唐利,何圣兵.二次通用旋转组合设计下磁絮凝工艺处理河道水效果[J].中国给水排水,2021,37(3):92.
YUAN Wen-jing,TANG Li,HE Sheng-bing.Effect of Magnetic Flocculation with Quadratic General Rotary Combination Design for Treatment of River Water[J].China Water & Wastewater,2021,37(14):92.
[7]孙军波,朱建文,胡惠宁,等.CPAM对河道泥水混合物的絮凝效果研究[J].中国给水排水,2021,37(11):84.
SUN Jun-bo,ZHU Jian-wen,HU Hui-ning,et al.Flocculation Effect of Cationic Polyacrylamide on Mud-water Mixture in River[J].China Water & Wastewater,2021,37(14):84.
[8]何嘉莉,袁耀芬,周沛良,等.自来水厂混凝剂自动精准投加系统建设与运行[J].中国给水排水,2021,37(18):139.
HE Jia-li,YUAN Yao-fen,ZHOU Pei-liang,et al.Construction and Operation of Automatic and Accurate Coagulant Dosing System in Waterworks[J].China Water & Wastewater,2021,37(14):139.
[9]张杨,李礼,佟怿维,等.低浊水中颗粒扰动对浊度测量的影响[J].中国给水排水,2022,38(20):131.
ZHANGYang,LILi,TONGYi-wei,et al.Influence of Particle Disturbances on Turbidity Measurements in Low Turbidity Water[J].China Water & Wastewater,2022,38(14):131.